10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available to be opened manually
as follows.
1Ensure if the rotor has stopped rotating.
z Listen carefully to ensure no rotating sound can be heard.
2Insert a screw driver into a hole to open door.
z Holes are located on the left and right sides of the unit.
z Insert a screw driver into the two holes and push forward to release the door.
11. Instructions for the rotor and tubes
CAUTION
z Read the instructions thoroughly, to properly load and use rotor.
z Do not exceed the allowable maximum speed of rotortube and adapters etc. Ensure the allowable
maximum speed of adapters is lower than the rotor’s maximum speed.
11.1 Rotor instructions
11.1.1 Rotor structure
Figure 11-1 The rotor structure
adapter mat adapter screw nut rubber cap
Rotor body mat Spindle hole