Page: 120
436-GC/456-GC
720 Detectors
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
FID Flame Ionization Detector
7
6
1 Signal Cable (391783000)
1
2 Signal Probe (200187300)
5 2a O-ring (baked, 0390596004)
2a
3
2
8
9
4a
4
10
11
12
13
14
15
16
3
Ignitor Cable (392548701)
4
Ignitor Probe (200187200)
4a O-ring (baked, 0390596004)
5 Screws, 8-32 x ¼” (1214200804)
6 Collector Tube (394958700)
7 Tower Screws, 8-32x2” ¾” (391866302)
8
Ceramic Insulator (2100003200)
9
Tower Top Plate (3400073100)
10
Detector Tower Body (100099300)
11
Flame Tip, 0.01 in. Capillary Tip (200187500)
11 Flame Tip, 0.02 in. Standard Tip (200193800)
12 Ferrule: Vespel (100%) max. 350°C (CR212200)
Graphite/Vespel 40/60 max. 400°C (CR213200)
Graphite (100%) max. 450°C (CR211200)
13
Aluminum Seal Washer (1500334701)
14
Detector Base (392547501)
15
Column Ferrule
16 Column Nut (394955100)
Temperature Probe (392537401)
Heater, 120V 125W- 392539601
230V 125W- 392539602
--- Flow Tube (200187600)
--- Large Stopper (394958600)