730 Electronic Flow Control
436-GC/456-GC Service Manual Revision B February 2019
Electronic Flow Control Type 25
The EFC25 is almost identical as the EFC21 the only exception is that the
pressure is monitored at the module itself rather than at the Injector. This allows
the EFC25 to be used with Gas or Liquid Sample Valve’s and Purge and Trap
devices upstream from the Injector.
Optimized use:
For Valved systems, Purge Trap systems, headspace
Flow & Pressure control
Use with split/splitless/PTV injectors
Liquid
Sample
* Valve
Pressure Control Unit
P -sensor
Pencil
Split
Line