Scion Instruments 436-GC - Removal and replacing FID;NPD Flame tip

Scion Instruments 436-GC
194 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page: 58
436-GC/456-GC
480 Procedures
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
Upgrade tool.
1.
Connect the Ethernet cable to the instrument and your laptop (Cross over cable).
2.
Switch on the GC.
3.
Start the Update tool.
4.
Type in the IP address of the instrument and “Update GC firmware”.
5.
Information will be send to GC.
6.
When completed reboot the 436-GC/456-GC to install the new firmware on the 436-
GC/456-GC.