436-GC/456-GC
510 Error codes
Page: 63
SCION Instruments
436-GC/456-GC Service Manual Revision B February 2019
Code
Error Description
Error level
Cause Condition(s)
GC action
Action needed
H2 flow set to 0.0 mL/min
80 Zone 6 Shorted Sensor Error Shorted sensor. Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (6)
Thermal Runaway (zone)
81 Column Oven Thermal Runaway Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
Check Column Oven zone
82 Zone 1 Thermal Runaway Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (1)
83 Zone 2 Thermal Runaway Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (2)
84 Zone 3 Thermal Runaway Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (3)
85 Zone 4 Thermal Runaway Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (4)
86 Zone 5 Thermal Runaway Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (5)
87 Zone 6 Thermal Runaway Error -
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (6)
Max Temper Exceeded (zone)
88 Column Oven Max Temper Exceeded Error Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check Column Oven zone
89 Zone 1 Max Temper Exceeded Error Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (1)
90 Zone 2 Max Temper Exceeded Error Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (2)
91 Zone 3 Max Temper Exceeded Error Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (3)
92 Zone 4 Max Temper Exceeded Error Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (4)
93 Zone 5 Max Temper Exceeded Error Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (5)
94 Zone 6 Max Temper Exceeded Error Defect Sensor
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (6)
Min Temper Exceeded (zone)
95 Column Oven Min Temper Exceeded Error
Turn device off. Temperature below -10 °C, only for
non callable devices
Electronics off (all)
H2 flow set to 0.0 mL/min
Check Column Oven zone
96 Zone 1 Min Temper Exceeded Error
Turn device off. Temperature below -10 °C, only for
non callable devices
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (1)
97 Zone 2 Min Temper Exceeded Error
Turn device off. Temperature below -10 °C, only for
non callable devices
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (2)
98 Zone 3 Min Temper Exceeded Error
Turn device off. Temperature below -10 °C, only for
non callable devices
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (3)
99 Zone 4 Min Temper Exceeded Error
Turn device off. Temperature below -10 °C, only for
non callable devices
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (4)
100 Zone 5 Min Temper Exceeded Error
Turn device off. Temperature below -10 °C, only for
non callable devices
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (5)
101 Zone 6 Min Temper Exceeded Error
Turn device off. Temperature below -10 °C, only for
non callable devices
Electronics off (all)
H2 flow set to 0.0 mL/min
Check zone (6)
Too Slow Response (zone)
109
Column Oven too slow response
Error
Sensor not used (wrong used), mismatch heated
Electronics off (all)
Check Column Oven zone