EasyManuals Logo

SciQuip Oven-S Series User Manual

SciQuip Oven-S Series
10 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #7 background imageLoading...
Page #7 background image
8
According to the diagrams shown above, In both cases of A and B, because of the power
limitation of heaters, it can only be heated up to 35 from 30 for the 5 mins of R1, still
5 lower than the "RET".
In Case A, it goes straight to the "H1"step,which includes another 5 mins to heat up to
40 , leaving the actual retaining time only 5 mins instead of the nominated, 10mins.
In case B, it will keep heating up to 39.9 (40 -0.1 ), and then the "Hold" step starts
for exactly 10 mins, which is 5 mins longer than that in Case A.
If it's the same situation in Segment 2, in Case A, the whole program runs 30 mins in total,
while 40 mins in Case B.
5. Press "Set" to enter the page of "Programme Parameter, Cycle", press
"Increase/Decrease" to set the times of repeat running of the whole program. The running
is to be repeated only if the segment setting ends with "Cycle" mentioned in 2.3.
6. Press "Set" again to finish setting of program control mode.
If a "Pause" of the running program needed, simply press "R/H/S", and
the "Run" indicator twinkles, and press it again to resume. To restart the
program, press the "R/H/S" and hold for 3 seconds to quit the existing
program first, the "Run" indicators goes off, and press "R/H/S" again to
restart the whole program.
Case B: HSD=0.1
Attention!
2
SETSET SETSET
SETSET SETSET
SETSET SETSET
SETSET SETSET
SETSET
Hold on for
3 seconds
Change/Confirm
button
Increase
button
Increase
button
Decrease
button
Change/Confirm
button
Increase
button
Decrease
button
Change/Confirm
button
Increase
button
Decrease
button
"Hold Time" (HT)
Change/Confirm
button
Increase
button
Decrease
button
Change/Confirm
button
Increase
button
Decrease
button
Change/Confirm
button
Increase
button
Decrease
button
Change/Confirm
button
Increase
button
Decrease
button
Change/Confirm
button
Increase
button
Decrease
button
The 10thprogram segment
Cycle time
Simplified instruction on the setting of program control time and temperature
9
Segment
"Ramp Time" (RT)
"Ramp End
Temperature" (RET)
Segment 1
Segment 1
"Segment 2
Ramp Time" (RT)
"Ramp End Temperature" (RET)
Segment 2
Segment 2
Hold Time (HT)
Segment 3
"Ramp Time" (RT)
"Ramp End Temperature" (RET)
Segment 3
SETSET
Increase
button
Decrease
button
Change/Confirm
button
Segment 3
SETSET
Change/Confirm
button
Segment 4
Ramp Time (RT)
SETSET
Increase
button
Decrease
button
Change/Confirm
button

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SciQuip Oven-S Series and is the answer not in the manual?

SciQuip Oven-S Series Specifications