Summary of a typical recalibration sequence:Summary of a typical recalibration sequence:
Action carried out by the user Result
The scales are switched off. No values or characters on the display.
Press the kg/lbs key and keep it depressed
while simultaneously pressing the start key.
The contents of the calibration counter is
shown on the display for 5 seconds
(flashing).
Release both keys. The contents of the calibration counter is
shown on the display for 5 seconds
(flashing).
Press the kg/lbs key for more than
1.5 seconds within the next 5 seconds.
The system switches to the calibration mode:
first "dec" appears on the display (flashing)
and then the current weight (0.0 - also
flashing).
5 seconds have passed and no key has
been pressed.
The scales are switched back to normal
weighing operation.
Place the test weight onto the scales. The measured weight of the test weight
appears on the display (flashing).
Decide whether to decrement or to
increment:
Incrementing:
Press the kg/lbs key for more than 1.5
seconds.
Decrementing:
No action required as the scales are already
in the dec mode.
The incrementing mode is activated; "inc"
appears on the display.
No output on the display.
Decrementing or incrementing:
Again briefly press the kg/lbs key (for less
than 1.5 seconds).
The measured weight is decreased or
increased by 10 g. Please remember that
pressing the key does not always change the
value on the display due to the low display
resolution. Only the rounded value is given.
Repeat this step until the measured weight
matches the test weight (format: xx.0 kg).
The display indicates the test weight (format:
xx.0 kg).
To correct this rounding error, decrement or
increment the value five times by another
10 g.
The display output does not change the first
four times.
After the fifth time the value has been
decremented or incremented, the measured
weight appears; it now matches the test
weight.
The display indicates the exact test weight
(format: xx.00 kg).
Press the kg/lbs key for at least 5 seconds
until the system is switched off.
The display does not show any characters or