seca page 1 (2) PC configuration program
17.06.03- Wint/Rei 30-34-00-672 b
Instructions for PC-based configuration
PC-based configuration is used for diagnosis or recalibration of all seca scales with modular electronics.
The following equipment is required:
PC with serial interface (RS -232)
seca software 'SERVA” 10-02-01-004
RS232 service set 68-60-00-002,
consisting of:
RS-232-service module 68-60-00-003
Connecting cable PC/service module 08-06-16-121
Modular cable 08-06-16-115
Bus cable for DETF-LEM 08-06-16-125
Loading weights
For scales with DETF sensor at least 5 weight stages up to the MAX
For scales with DMS load cell at least 1 weight stage (=MAX)
Software Installation
If your 'SERVA'-software is in a ZIP-file, it has to be unpacked at first.
Execute the program's SETUP.EXE and follow the instructions displayed by the setup program.
Connecting and calibrating a scale
If the service module has not been connected to the PC yet, shut the PC down, connect the service
module to the serial interface and reboot the PC.
PC
Drucker
Printer
Vorderseite / Frontside Rückseite / Backside
Waage \ Scale
Anzeige / Display
Switch off the scale!
For calibration, a connection between the service module and the measured-value
acquisition module of the scale to be calibrated must be established. For this purpose, a
modular cable must be connected to the scale’s SeSAM bus.