seca page 1 (2) PC configuration program
17.06.03- Wint/Rei 30-34-00-672 b
Instructions for PC-based configuration
PC-based configuration is used for diagnosis or recalibration of all seca scales with modular electronics.
The following equipment is required:
• PC with serial interface (RS -232)
• seca software 'SERVA” 10-02-01-004
• RS232 service set 68-60-00-002,
consisting of:
• RS-232-service module 68-60-00-003
• Connecting cable PC/service module 08-06-16-121
• Modular cable 08-06-16-115
• Bus cable for DETF-LEM 08-06-16-125
• Loading weights
• For scales with DETF sensor at least 5 weight stages up to the MAX
• For scales with DMS load cell at least 1 weight stage (=MAX)
Software Installation
If your 'SERVA'-software is in a ZIP-file, it has to be unpacked at first.
Execute the program's SETUP.EXE and follow the instructions displayed by the setup program.
Connecting and calibrating a scale
• If the service module has not been connected to the PC yet, shut the PC down, connect the service
module to the serial interface and reboot the PC.
Vorderseite / Frontside Rückseite / Backside
Waage \ Scale
Anzeige / Display
• Switch off the scale!
• For calibration, a connection between the service module and the measured-value
acquisition module of the scale to be calibrated must be established. For this purpose, a
modular cable must be connected to the scale’s SeSAM bus.