EasyManua.ls Logo

Seko PH - Dosing Operation Examples; Troubleshooting; Troubleshooting Common Issues

Default Icon
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0000136965
Rev. 1.1
8
3.5__Dosing and adjustment examples
Dosing Operation/Pause (ON
-
OFF)
The instrument permits automatic control and modification of
the chemical measurement through Set Point 1, adjusting
dosing by m
eans of On
-
Off Relay 1.
The dosing example on the side can be set using the following
parameters:
Set Point 1 = 7.20 pH
Type of Dosing = Alkaline
Hysteresis = 0.40
The instrument will begin dosing for values below 7.20 and
will stop dosing when above 7.60.
If Acid is set for Type of Dosing, dosing will be the exact opposite: dosing will
begin above 7.20 and will stop below 6.80 pH. The instrument permits two independent dosings by setting Set
Point 2 associated with Relay 2.
Dosing with delays
The instr
ument permits dosing start and end delays. The dosing
example on the side can be set using the following parameters:
Delay Start (D
-
1) = 5 minutes
Delay End (D
-
2) = 5 minutes
This function is available for all of the previously described
settings: ON
-
OFF,
Timed, Prop.
Note: The variables shown above are independently available for
Set Point 1 and Set Point 2.
4
TROUBLESHOOTING
The device does not come on…
o
Make sure that the power cables are correctly connected
o
Check whether the power supply is pre
sent
The display does not light up…
o
Regulate the display contrast
The chemical measurement function does not work…
o
Check the probe connection
o
Calibrate as instructed in the manual
o
Replace the probe
The mA output does not change…
o
Check the cable connection
o
Use the “Manual Control” function in the main menu to check whether the output produces the
desired effect.
o
Check the electrical characteristics of the remote device (maximum load: 500 ohm).
The relays do not work…
o
Check whether the instrument is correctly
powered
o
Check the settings in the main menu
Note:
If the anomaly persists, please contact the supplier.

Related product manuals