Redox Probe Calibration
Wash
Keep probe into Buffer solution
G1><A.1;</4!
Press Cal Key 3 Seconds
Set Redox calibration
5
*,%'(###-.3::#D1>
Calibration During 1 minutes
K1<;########,&:##
6
*,%'(#L51><;=#M&&N!
Quality Probe
7
Press Cal Key 3 Second
10
Normal Status
11
enter
cal
enter
cal