-13-
Push CAL key.③
Stability of electrode potential
is discriminated, then calibrates
it to the pH value of standard
solution at that temperature
of discrimination.
Push CAL key again even if you
calibrate with the same standard solutions.
It is possible to calibrate many times
with the same standard solutions, but
the only final data is effective.
Immerse the electrode in standard solution of pH 4 .④
Push CAL key.⑤
Stability of electrode potential
is discriminated, then calibrates
it to the pH value of standard
solution at that temperature
of discrimination.
Measured value stops flashing;
the type of used standard solution
is shown.
Push CAL key again even if you calibrate
with the same standard solutions.
It is possible to calibrate many times
with the s ame standard solutions, but
the only final data is effective.
『』,Set Value Display shows
which means these types were used to complete the calibrations.