Place a maximum of 20 ml of normal
saline or prescribed medication in the
nebulizer cup.
Connect yellow tubing quick-connect
ing to nebulizer cup.
Connect clear tubing quick-connect
ing to the cap.
Press red tubing onto connector.
Connect blue corrugated tubing to
exhalation port.
Connect the tubing to the
corresponding-colored connectors.
Do not obstruct green remote port
Ensure Master Switch is OFF.
Then, connect the IPV® to a 50 psi
gas source.
Assembling Phasitron
Patient Connections
IPV
®
-1C Quick Reference Guide
Mouthpiece
Instruct patient to keep
lips and cheeks splinted
to ensure the pulses are
directed into the lungs.
Mask
It is recommended
to use a resuscitation
mask or similar non-
vented mask.
Direct
The Phasitron can be
aached directly to
an articial airway or
airway adaptor.
Before you start
NOTE: The IPV®-1C is not
recommended for use in
neonates. For patients under ve
years of age without an articial
airway, mask conguration is
recommended