■
Pulse Flow Mode Operation
• Pulse Flow Mode should only be used under the direction of a qualified
clinician or physician.
• When operating in Pulse Flow Mode, an alarm will beep after 15 seconds
if inspiration is not detected. If another 15 seconds elapses and no
inspiration is detected, the unit will switch to Continuous Flow Mode
at the last Continuous Flow Mode rate setting used.
WARNING Pulse Flow settings should be determined for each patient individually.
Settings from Continuous Flow applications may not be applicable to
Pulse Flow Mode.
WARNING When operating in Pulse Flow Mode, to ensure adequate breath
detection, the length of the tubing/cannula must not exceed
7 feet (2.1m).
WARNING As with all conserving devices, the ECLIPSE may not be able to detect
some respiratory efforts.
STEP 5: ADJUST FLOW RATE TO PRESCRIBED LEVEL
■
Using the Flow Rate Setting buttons, adjust the flow rate to
the prescribed setting supplied by your physician.
WARNING DO NOT change from prescribed flow rate setting without consulting
with a qualifi
ed clinician or physician.
28
|
ECLIPSE
TM
Model 1000 User Manual