EasyManua.ls Logo

Sharp 21F-PA18 - Purity Adjustment

Sharp 21F-PA18
89 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21F-PD250 / 21F-PT220 / 21F-PA18 / 21F-PA18(B)
3 – 15
2. PURITY ADJUSTMENT
No.
Adjustment point
Adjustment procedure/conditions Waveform and others
1. Receive the GREEN-ONLY signal. Adjust the
beam current to ~700 µA .
2. Degauss the CRT enough with the degausing coil.
Note: Follow the Job Instruction Sheet to adjust
the magnetic field.
(Reference: page 3-5)
3. Maintain the purity magnet at the zero magnetic
field and keep the static convergence roughly
adjusted.
4. Observe the points a, b,as shown in Fig. 1-1
through the microscope. Adjust the landings to A
rank requirement.
5. Orient the raster rotation to 0 eastward.
6. Tighten up the deflection coil screws.
Tightening torque: 108 ± 20 N (11kgf ± 2 kgf)
7. Make sure the CRT corners landing meet the A
rank requirements. If not, stick the magnet sheet
to correct it.
Note: This adjustment must be done after
warming up the unit for 30 minutes or
longer with a beam current over 700
µ
A.
Note:
* For the following colours press R/C RGB(Hex 7E)
key to change.
PURITY ADJ.
(No need if used
ITC CRT)
1
a
b
A
B
A
30mm 30mm
B
A=B
A=B
Rank "A"
(on the right of the CRT)
Rank "A"
(on the left of the CRT)
Fig. 1-1
Fig. 1-2
Fig. 1-3
* Press R/C RGB key for 1 sec-
ond in NORMAL MODE, the
colour will change to RGB
mono colour mode.
Signal-colour
screen cleared
GREEN
ONLY
BLUE
ONLY
RED
ONLY
Set to service mode by remote con-
troller then press factory process R/C
RGB key to change to RGB mono col-
our mode.

Table of Contents

Related product manuals