SLOW RESPONSE Free Water in System Check drain & inspection ports in the main system
& sample pipe work.
Wrongly Sized or Refer to Section 3.2 of manual. & ensure sample
Incorrect Material on tube material & flow rates are as
Sample Lines recommended.
DRY READING Autocal Wrongly Set Refer to Section 7.0 of manual & re-check Autocal.
If this does not cure problem return for
calibration.
Comparison with Refer to Section 7.0 of manual & re-check Autocal.
Secondary Standard If this does not cure problem return for calibration.
Instrument or Calibration
Gas
CONSTANT READING Condensation in Sample Condensation will occur if temperature of the
System sample system, or main pipe, at any point is below
(colder) than the DP temperature of the sample
gas. Once having formed, the sample reaching the
sensor will have a DP equal to temperature of
the condensate, regardless of the DP at the sample
point.