23 | Page
INSTALLATION
ACCESS PORT STOPPER
Each incubator ships with a silicone stopper installed in the access port located
on the right side of the incubation chamber interior.
• The stopper should always be installed inside the chamber to obtain the
best temperature uniformity and prevent condensation from forming
inside the port.
o Do not install the stopper on the outside of the incubator.
• Wires for thermocouples and other sensor probes may be introduced into
the chamber through the access port. The stopper may be put in place
over the wires.
• The unit will not meet its temperature or CO
2
specifications without the
stopper installed.