EasyManua.ls Logo

Shimadzu UV-2550 - User Manual

Shimadzu UV-2550
67 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
206-94649D
Sep. 2008
INSTRUCTION MANUAL
User's System Guide
SPECTROPHOTOMETER
UV-2450
UV-2550
Read the instruction manual thoroughly before you use the product.
Keep this instruction manual for future reference.

Questions and Answers

  • J
    John IrwinAug 8, 2025
    How to fix Shimadzu UV-2550 Measuring Instruments light source?
    • K
      Kevin LevyAug 8, 2025
      If the light sources in your Shimadzu Measuring Instruments aren't lit, turn off the photometer unit and computer. Restart the startup procedure, watching to see if the lamps light up. If they don't, the lamps likely need replacement. The Wl lamp typically lasts for 2000 hours, and the deuterium lamp for 500 hours of operation. Replace the failed lamp and begin the startup procedure again.
  • M
    Melissa WoodwardAug 23, 2025
    What to do if something is blocking light beams in Shimadzu UV-2550 sample compartment?
    • J
      Joshua BanksAug 23, 2025
      If you find any material blocking the light beams in the sample compartment of your Shimadzu Measuring Instruments, remove the blockage and reinitialize the system.

Summary

Precautions for Safe Operation

Emergency and Power Outage Procedures

Emergency Operation

Procedure for turning off the instrument in an emergency situation.

Operation at Power Outage

Steps to follow during an electrical failure to ensure proper instrument handling.

Equipment Warning Labels

Hot Surface Warning

Warning about hot surfaces and burn risk before lamp replacement.

Electrical Safety Warning

Risk of Electric Shock

Warning about electric shock risk when changing fuses or voltage.

Chapter 1 Inspection

Chapter 2 Installation

Chapter 4 Maintenance and Checks

Troubleshooting Common Issues

Solutions for initialization failures, scanning problems, and abnormal readings.

Light Source Replacement Guide

Procedure for safely replacing the WI and D2 lamps.

Fuse Replacement Instructions

How to replace fuses for the power supply, with safety warnings.

Chapter 5 Technical Specifications

UV-2450;2550 Performance Specs

Detailed specifications including wavelength range, resolution, and accuracy.

Shimadzu UV-2550 Specifications

General IconGeneral
BrandShimadzu
ModelUV-2550
CategoryMeasuring Instruments
LanguageEnglish