380002M 21
Semi-micro and Analytical Balance
LF Series
Operation Manual
To ensure safe and proper use of the balance, please read
this manual carefully.
After reading this manual, store it in a safe place near the
balance, so you can review it as needed.
SHINKO DENSHI CO., LTD
This document hosted by: www.oldwillknottscales.com

Table of Contents

Questions and Answers

  • M
    Marissa McknightAug 7, 2025
    What to do if the displayed weight is incorrect on Shinko Denshi Scales?
    • E
      ericabenderAug 7, 2025
      If the displayed weight is incorrect, there could be several reasons: * The balance hasn't been used for a while or has been moved. In this case, perform span adjustment. * The adjusters are not flat, and the balance isn't horizontal. Ensure the balance is horizontal. * The tare range is either set or not set correctly. Remove the sample and zero the readout by pressing the [Zero/Tare] key to continue.
  • K
    Kayla PetersonAug 11, 2025
    Why is nothing displayed when I turn on my Shinko Denshi LF Scales?
    • J
      Jacob RiosAug 11, 2025
      If nothing is displayed when you power on the balance, it could be due to the following cause: * The AC adapter may not be properly connected. Ensure the AC adapter is securely connected. * If the adapter is connected, the electric components of the balance or the AC adapter may have failed. Replace the AC adapter with a properly working adapter of the same model to see if that resolves the problem.
  • J
    Joshua SingletonAug 16, 2025
    What to do if the weight indication contains an error on Shinko Denshi Scales even after calibration?
    • T
      Tammy LivingstonAug 16, 2025
      If the weight indication still contains an error after calibration, consider these potential causes: * External influences like wind or vibration during calibration. Ensure the balance is free from such influences during calibration. * The calibration weight differs slightly from the checking weight. Use the same weight for both calibration and checking.
  • R
    Richard HawkinsAug 20, 2025
    How to fix flickering display on Shinko Denshi LF?
    • D
      Daniel BrownAug 20, 2025
      If the display flickers, it may be due to external influences such as wind and vibration. Increase the setting values of relevant functions.
  • A
    Amber KingAug 25, 2025
    What to do if I am unable to output anything from Shinko Denshi LF Scales?
    • J
      Jennifer PorterAug 25, 2025
      If you're unable to get any output, consider the following: * The output function settings might not be configured correctly. Ensure the balance’s communication conditions match those of the external device. * Communication settings may not match the external device. Check output by matching the balance’s communication conditions with the external device. * The cable may be wrongly connected. A crossover cable might be required to connect the balance to a common PC. * The cable might be disconnected or not properly connected. Ensure the cable is properly connected.
  • A
    Ariel SmithAug 29, 2025
    Why does the 'M' continuously flash on Shinko Denshi Scales?
    • A
      Ashley MccallAug 29, 2025
      The 'M' may continuously flash due to external influences such as wind and vibration. Take remedial actions or relocate the balance.
  • J
    joeblackwellSep 2, 2025
    What to do if Shinko Denshi Laboratory Equipment shows 'o-Err'?
    • A
      Andrew PorterSep 2, 2025
      If your Shinko Denshi Laboratory Equipment displays 'o-Err', it could be due to several reasons: The weight of the sample exceeds the balance's capacity, in which case you should unload the sample and weigh it in parts. It may also be that the digits in the addition or calculation result went beyond the display limit, so clear the addition result and try again. Another possibility is that the unit conversion coefficient is too small, so set a larger coefficient. If the error persists even without any load on the weighing pan, there may be a mechanical failure.

Shinko Denshi LF Series Specifications

General IconGeneral
BrandShinko Denshi
ModelLF Series