Questions and Answers

  • R
    Robert HouseAug 11, 2025
    What to do if '[o-Err]' appears on Shinko Denshi LN 223CE before capacity is reached?
    • B
      bryantcarlAug 11, 2025
      If the Shinko Denshi Scales display shows '[o-Err]' before reaching capacity, it could be due to the total weight, including the container, exceeding the measurement range. Check the total weight. Alternatively, the mechanism section might be damaged, so review the container.
  • J
    jessicaacostaAug 17, 2025
    Why is the display turned off on my Shinko Denshi Scales when using batteries?
    • K
      Kevin ShawAug 17, 2025
      If your Shinko Denshi Scales display is off when running on batteries and the battery icon blinks, or there is no display, it could be due to the auto power off function. Turn the power back on or disable the auto power off function. Alternatively, the batteries may be exhausted, so recharge them. As another option, you can operate the balance on the AC adapter.
  • M
    margaretwilliamsAug 22, 2025
    Why does the weight indication contain an error on Shinko Denshi LN 223CE?
    • M
      munozleslieAug 22, 2025
      If the Shinko Denshi Scales show a weight indication error, ensure the tare range is set correctly. Also, check that the balance is kept horizontal. If the balance hasn't been used for a while or has been moved, perform a span adjustment.
  • B
    Brian HamiltonAug 29, 2025
    What does '[u-Err]' mean on Shinko Denshi LN 223CE Scales and how to fix it?
    • T
      Todd WheelerAug 29, 2025
      If the Shinko Denshi Scales display shows '[u-Err]', a foreign object might be trapped between the pan (or pan base) and the balance. Remove the pan and inspect the surface underneath. The mechanism section may also be damaged.
  • S
    samantha39Sep 4, 2025
    How to fix no display on Shinko Denshi LN 223CE?
    • G
      Geoffrey OconnorSep 4, 2025
      If your Shinko Denshi Scales have no display, ensure the AC adapter is properly connected. If it is, the batteries may be exhausted, in which case, recharge them.