00 User Manual
28
|
P a g e
Alarm: Lower limit of a
flow rate
0x009A ~ 0x009B
Write
Y
Read
Y
Description
Set alarm value for a lower flow rate limit
Value type
U
I
NT 32
Notes
Alarm values = Value (0x009A)*65536 + Value (0x009B)
When the flow rate is below a set value, an alarm will be triggered.
Change password 0x00AE ~ 0x00AF
Write
Y
Read
Y
Description
Change the default password
Value type
U
I
NT 32
Notes
Password values = Value (0x00AE)*65536 + Value (0x00AF)
Available: 00000 ~ 99999
Offset calibration 0x00F0
W
rite
Y
Read
N
Description
Reset or calibrate the offset
Value type
U
I
NT 16,
Fixed
value
0xAA55
Notes
To reset or calibrate the offset, write 0xAA55 to register 0x00
F0
.
When you execute this function, make sure there is NO flow in the flow
channel.
Reset accumulated
flow rate
0x00F2
Write
Y
Read
N
Description
Reset the accumulated or totalized flow rate value
Value type
U
I
NT 16, Fixed value 0x0001
Notes
To reset the accumulated or totalized flow rate value, write 0x0001 to register
0x00F2.
Write protection 0x00FF
Write
Y
Read
N
Description
Write protection disabler for a set value to a specific register.
Value type
U
I
NT 16,
Fixed
value
0xAA55
Notes
This function is
en
abled at the time of product shipment. To enable the write
function of a specific parameter, such as GCF, the user needs to send 0xAA55
to the register 0x00FF, and then the write function will be enabled (write
protection is disabled). After the write execution is completed, the firmware
will automatically re-enable the write protection.