EasyManua.ls Logo

Siemens 7MF0400 - 8.2.5 Absolute pressure from gauge and differential pressure series

Siemens 7MF0400
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step response time T
63
(without electrical damping) - gauge pressure from differential pressure series
20 mbar/2 kPa/8.031 inH
2
O Approx. 0.160 s
60 mbar/6 kPa/24.09 inH2O Approx. 0.150 s
250 mbar/25 kPa/3.6 psi
600 mbar/60 kPa/240.9 inH
2
O
1600 mbar/160 kPa/642.4 inH
2
O
5000 mbar/500 kPa/2008 inH
2
O
30 bar/3 MPa/435 psi
Approx. 0.135 s
100 bar/10 MPa/1450.3 psi Approx. 0.145 s
Effect of mounting position - gauge pressure from differential pressure series
≤ 0.7 mbar/0.007 kPa/0.01015266 psi per 10° incline
(correct the zero point with position error compensation)
8.2.5 Absolute pressure from gauge and differential pressure series
Conformity error at limit point setting, including hysteresis and repeatability - absolute pressure from gauge and differential
pressure series
Measuring span ratio r (spread, Turn-Down) r = max. measuring span/set measuring span and nominal
measuring range
Linear characteristic curve
r ≤ 10 10 < r ≤ 30
All measuring cells ≤ 0.1% ≤ 0.2%
Effect of ambient temperature - absolute pressure from gauge and differential pressure series
In percent per 28 °C (50 °F)
250 mbar/25 kPa/3.6 psi ≤ (0.15 • r + 0.1)
1300 mbar a/130 kPa a/18.8 psi a
5 bar a/500 kPa a/72.5 psi a
30 bar a/3000 kPa a/435 psi a
100 bar a/10 MPa a/1450.3 psi a
160 bar a/16 MPa a/2321 psi a
400 bar a/40 MPa a/5802 psi a
700 bar a/70 MPa a/10152.6 psi a
≤ (0.08 • r + 0.16)
Long-term stability at ±30 °C (±54 °F) - absolute pressure from gauge and differential pressure series
In 5 years ≤ (0.25 • r)%
Technical data
8.2 Measuring accuracy of SITRANS P320
SITRANS P320/P420 with 4 to 20 mA/HART
72 Compact Operating Instructions, 09/2018, A5E38874562-AB

Table of Contents