75 GB/IE
Guidelines and manufacturer‘s declaration -
electromagnetic interference
The SBM 68 is intended for operation in an electromagnetic en-
vironment such as is described below. The client or the user of
the SBM 68 should ensure that is used in this type of environ-
ment.
Immunity
tests
IEC 60601-
test level
Compli-
ance level
Electromag-
netic envi-
ronment
- guideline
Electrostatic
discharge
(ESD) in ac-
cordance
with IEC
61000-4-2
± 6 kV
Contact
discharge
±8 kV air
discharge
± 6 kV
Contact
discharge
±8 kV air
discharge
Floors should
be made from
wood or con-
crete or be de-
corated with
ceramic tiles. If
the floor is co-
vered with syn-
thetic material,
the relative hu-
midity should
be at least
30%.
Magnetic
field for
supply
frequency
(50/60 Hz)
In accordance
with IEC
61000-4-8
3 A/m 3 A/m The quality of
the supply fre-
quency should
correspond to
that of a typi-
cal commercial
or hospital
environment.