5.
Editing Test Parameters
Default test type parameters are shown below.
There are a total of 8 dierent Line condions that may be congured. They are displayed
in the following table:
AC Hipot Mode
• Voltage: 1.24 kV
• HI-Limit: 10.00 mA
• Lo-Limit: 0.00 mA
• Ramp: 0.1 s
• Dwell: 1.0 s
• Frequency: 60 Hz
• Arc Detect: OFF
• Arc Sense: 5
• DUT Output: On
• Connect: O
Funconal Run
• Volt-Hi: 125.0 V
• Volt-Lo: 0.0 V
• Amp-Hi: 10.0 A
• Amp-Lo: 0.0 A
• Power-Hi: 1000 W
• Power-Lo: 0 W
• PF-Hi: 1.000
• PF-Lo: 0.000
• Leakage-Hi: 10.00 mA
• Leakage-Lo: 0.00 mA
• Delay: 0.2 s
• Dwell: 1.0 s
• Connect: O
DC Hipot Mode
• Voltage: 1.50 kV
• HI-Limit: 5.00 mA
• Lo-Limit: 0.00 mA
• Ramp: 0.1 s
• Dwell: 1.0 s
• Arc Detect: OFF
• Arc Sense: 5
• DUT Output: On
• Connect: O
Line Leakage Mode
• Volt-Hi: 125.0 V
• Volt-Lo: 0.0 V
• Leakage-Hi: 6000 µA
• Leakage-Lo: 0.00 µA
• Line: 7
• MD: MD A
• Probe: G-L
• Delay: 1.0 s
• Connect: OFF
Insulaon Resistance
• Voltage: 500 V
• HI-Limit: 0 mΩ
• Lo-Limit: 1 mΩ
• Delay: 1.0 s
• DUT Output: On
• Connect: O
Ground Bond Test
• Current: 30 Amps
• HI-Limit: 100 mΩ
• Lo-Limit: 0 mΩ
• Dwell: 1.0 s
• Oset: 0 mΩ
• Frequency: 60 Hz
• Connect: O
Line 1 2 3 4 5 6 7 8
Neutral OPEN OPEN OPEN OPEN CLOSE CLOSE CLOSE CLOSE
Reverse OFF ON OFF ON OFF ON OFF ON
Ground OPEN OPEN CLOSE CLOSE OPEN OPEN CLOSE CLOSE