SOMATICS, LLC REV: 3.1 09/15/2015
20
5.8 Following items are not adjustable and their failure requires factory service:
Stimulus duration (pulse count)
Pulse width
Pulse frequency
fig. 5.2
5.8 FACTORY CONFIGURATION
In normal use these presets need not to be changed unless the system is moved to a
different country using different line frequency or date display convention or language.
To change any of the above items enter the calibration mode using jumper 3. Note that
the display should show CALIBR. Now turn the Flexdial clockwise one click and the
display should show CONFIG.
Press the Flexdial button and rotate the Flexdial to select 50 or 60 Hz notch filter
depending on the frequency of the power mains.
Press the Flexdial button again and rotate the Flexdial to select the date format; either
Day/Month/Year or Month/Day/Year.
Press the Flexdial button again and rotate the Flexdial to and select 24 hour or 12 hour
format.
Press the Flexdial button again and rotate the Flexdial to select from the available list of
languages which will be used to print out the report.
Press the Flexdial button again and CONFIG will be displayed. All the configuration
settings are stores in the battery backed memory and will remain there even with the
power switched off.