7. Cleaning, maintenance, disinfection and sterilization
Equipment
The equipment should only be treated with non-inflammable and non-explosive cleaning and
disinfection agents. Care should be taken that no liquids soak into the equipment.
ATTENTION: The following points are only valid for Söring products!!!
Handpieces
Handpieces as well as their cables and connectors should be treated with non-inflammable and
non-explosive cleaning as well as disinfection agents. They are autoclavable up to 134°C.
Reusable patient electrode
The silicon electrode including cable and connector are autoclavable up to 134°C.
Footswitches (IPX8/AP)
Footswitches may be treated with all common cleaning and disinfection agents but may not be
autoclaved.