90207/90217 ABP Monitors
1-16
Monitor Report Condition
EC07
EC78 Clogged luer filter
EC08
EC18 Too few data entries to accurately determine blood pressure. The
message may indicate that the cuff is not being worn by the patient (taken
off but left connected to the monitor). The message may also indicate that
motion artifacts cause the majority of the incomplete data.
EC28 Diastole above 200 mmHg
EC38 Pulse pressure less than 16 mmHg
EC48 A) Movement artifact at mean arterial pressure
B) Heart rate arrhythmia
EC58 A) Movement artifact at diastole
B) Heart rate arrhythmia
EC68 Division by zero
EC09
EC19 Contradictory instructions sent to hardware (e.g., pump on and valve
open)
EC29 Diastolic pressure value cannot be obtained from the data available.
EC39 90207
Systolic pressure cannot be obtained from the data available.
90217
Algorithm could not process input data quickly enough resulting in an
input queue overflow.
EC49 90207
Mean arterial pressure cannot be obtained from the data available.
90217
This monitor must be initialized.
EC59 Heart rate value cannot be obtained from the data available.
EC69 Heart rate value cannot be obtained from the data available.
EC79 Bleed steps were too small. This may be caused by a partially obstructed
air hose. All blood pressure attempts following this message are inhibited.
Attempts can be enabled by turning the power switch off then on.
EC99 Unexpected or contradictory data (such as a negative cuff pressure).
LLL
EC16 Low battery detected prior to start of measurement.
EC26 Low battery detected after measurement started. Usually caused by the
pump drawing enough current to lower the battery voltage.
Lbb The report does not print an event code for this condition, which is a low
backup battery. Contact Spacelabs Medical for replacement of the battery.