Ambulatory Blood Pressure
Monitors
90207/90217
Operations Manual
070-0137-03 Rev. F

Questions and Answers

  • D
    Donna BradleyAug 17, 2025
    What to do if Spacelabs 90207 display is incorrect?
    • R
      rick78Aug 17, 2025
      If the display is incorrect, it may be due to several reasons. First, the data may not be retained, in which case you should replace the backup battery. Second, there might be low or no power, so check the batteries and replace or recharge them as needed. Also, the issue might be a time-out, an air leak, improper cuff size, or the cuff not being properly attached. Try to isolate the cause and correct it.
  • F
    fturnerAug 22, 2025
    Why does my Spacelabs Blood Pressure Monitor (90207 only) display event code “Lbb” during selftesting?
    • C
      Chelsea ToddAug 22, 2025
      The monitor displays event code “Lbb” during self-testing because of a low backup battery condition. Replace the backup battery before continuing.
  • F
    fgregoryAug 25, 2025
    Why is my Spacelabs 90207 Blood Pressure Monitor display incorrect?
    • R
      Robert AllenAug 26, 2025
      An incorrect display on your Spacelabs Blood Pressure Monitor may be due to several reasons. It could be caused by data not being retained, in which case you should replace the backup battery. Another cause can be low or no power; check the batteries for a full charge and replace or recharge them if needed. Also, the issue might be related to a time-out, an air leak in the system, using the wrong cuff size, or the cuff not being properly attached. Try to isolate the cause and correct it.
  • K
    Kristen BrooksAug 28, 2025
    What to do if Spacelabs Blood Pressure Monitor displays 'LLL' and alarm sounds?
    • A
      Anna GregoryAug 28, 2025
      If your Spacelabs Blood Pressure Monitor displays 'LLL' and an alarm sounds, this indicates a low main battery condition. Immediately turn off the monitor and replace the batteries within 60 seconds to continue monitoring.
  • S
    susanquinnAug 30, 2025
    What does event code Lbb mean on Spacelabs 90207?
    • W
      wendyriveraAug 30, 2025
      If your Spacelabs Blood Pressure Monitor (90207 only) displays event code Lbb during self-testing, it indicates a low backup battery condition. Replace the backup battery before continuing.
  • K
    kristennguyenSep 2, 2025
    What to do if my Spacelabs 90207 Blood Pressure Monitor cuff is too tight?
    • L
      Lacey KennedySep 2, 2025
      If the cuff is too tight, it could be because the cuff was placed on the patient too tightly. Reposition the cuff. Another possible cause is the air pump staying on too long, in which case you should return the monitor to Spacelabs Healthcare for service.
  • D
    deborah86Sep 4, 2025
    What to do if my Spacelabs Blood Pressure Monitor cuff is too loose?
    • K
      Karen BlackburnSep 5, 2025
      If the cuff is too loose, it might be because it was placed on the patient too loosely. Reposition the cuff. Another possible cause is the air pump not staying on long enough, in which case you should return the monitor to Spacelabs Healthcare for service.
  • E
    Elizabeth FitzgeraldSep 7, 2025
    How to fix incorrect time displayed on Spacelabs Blood Pressure Monitor?
    • J
      Justin GardnerSep 7, 2025
      The time displayed may be incorrect because the batteries were removed for more than one minute. Re-initialize the monitor.
  • L
    Lori RomeroSep 10, 2025
    What to do if the Spacelabs Blood Pressure Monitor cuff is too tight?
    • T
      Toni EspinozaSep 10, 2025
      If the cuff feels too tight when using your Spacelabs Blood Pressure Monitor, it might be because the cuff was placed on the patient too tightly; try repositioning it. If the air pump stays on for too long, return the unit to Spacelabs Medical for service.
  • P
    paigemartinSep 10, 2025
    Why does my Spacelabs 90207 display “LLL” and alarm sound?
    • L
      Leslie PorterSep 12, 2025
      The monitor displays 'LLL' and the alarm sounds because of a low main battery condition. Immediately power OFF the monitor and replace the batteries within 60 seconds to continue monitoring.

Summary

Operation

Overview and Monitor Features

General description of ABP monitors, their features, and connection modes.

Battery Replacement Procedures

Instructions for replacing main and backup batteries in the ABP monitors.

Monitor Initialization

Steps for initializing the ABP monitors via PC Interface, Base Station, or Report Generator.

Modem Connection Setup

Details modem speed compatibility and connection procedures for ABP monitors.

Office Check Mode Operation

Explanation of the monitor's office check mode for initial performance verification.

Patient Preparation and Precautions

Guidelines for patient preparation and essential precautions for using the ABP monitor.

Data Transfer and Cleaning

Procedures for transferring patient data and cleaning the ABP monitor, cuff, and pouch.

Event Code Interpretation

Comprehensive list and explanation of event codes displayed by the ABP monitors.

Accuracy

Checking Monitor Accuracy

Procedure for annually checking the accuracy of the ABP monitor's pressure readings.

Accuracy Test Requirements

List of required equipment and step-by-step instructions for the accuracy check.

Troubleshooting and Servicing

Problem Solving Checklist

A checklist to identify and resolve common problems encountered with the ABP monitors.

Servicing and Warranty

Information regarding product warranty and procedures for returning the monitor for service.

Symbols and Abbreviations

International and Safety Symbols

Lists and describes international and safety symbols used on Spacelabs Medical products.

Electrical and Interface Symbols

Explains various electrical, power, and interface symbols found on the equipment.

Control and Data Symbols

Lists symbols related to interface controls, data input/output, and menus.

Equipment and Environmental Symbols

Describes symbols indicating equipment type, handling, and environmental conditions.

Connection and Indicator Symbols

Explains symbols for various connections, alarms, and indicators on the device.

Functional and Warning Symbols

Details symbols for audio, warnings, connections, and status indicators.

Abbreviations as Symbols

Lists abbreviations that are used as symbols in the manual and on the equipment.

Spacelabs 90207 Specifications

General IconGeneral
BrandSpacelabs
Model90207
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals