Safety
Instructions for use • OR1™ Table Interface • ZWV442_EN_V1.0_02-2021_IFU_CE-MDR 9
Observe the information on electromagnetic compatibility; see chapter Electromagnetic
compatibility [p.22].
In case of uncertainties, seek expert advice from KARLSTORZ.
Before use, a clinical/biomedical engineer or an EMC specialist should carry out an ad-hoc
test of the electromagnetic radiation.
The use of accessories and cables other than those specified in the instructions for use may
result in increased emissions or decreased immunity of the product. When using other
accessories and cables, the operator is responsible for checking compliance with
IEC60601-1-2 for this particular product.
To prevent increased electromagnetic emissions or reduced electromagnetic immunity of
the product, only use accessories, transducers, and cables recommended or supplied by
the manufacturer.
3.8 Observing ambient conditions
If the device is operated in an environment which is not suitable, patients, users and third
parties may be injured.
Always operate the product in the prescribed ambient conditions.
When using explosive narcotic gases: Operate the product outside of the hazard zone.
Do not use the product in the presence of flammable anesthetics.
The product must not be operated in oxygenated environments.