Strasbaugh Software Calibration Procedures
Version 4.0 - February 1998 10 - 121
SLURRY FLOW 2 CALIBRATION
Repeat as for slurry flow 1, making the following substitutions to duplicate the
slurry flow calibration procedures for slurry flow 2:
1. On the CALIBRATION menu:
20 = SLURRY 2 FLOW FACTOR variable
2. In the text of the procedure, substitute slurry 2 for slurry 1.
PUMP 3 CALIBRATION
Repeat as for slurry flow 1, making the following substitutions to duplicate the
slurry flow calibration procedures for slurry flow 2:
1. On the CALIBRATION menu:
75 = PUMP 3 FLOW FACTOR variable
2. In the text of the procedure, substitute pump 3 for slurry 1.