Strasbaugh Software Calibration Procedures
Version 4.0 - February 1998 10 - 167
B. COMPARE THE ACTUAL TEMPERATURE WITH THE SCREEN
STATUS AREA DISPLAY
1. Compare the Actual Temperature (from the temperature control unit
display) with the screen status area display for Cooling Water In:
GOAL: The Actual Temperature must be within +/- 2 % of the
COOLING WATER TEMPERATURE IN status screen
display.
If the goal is met, no adjustment is necessary. Go on to the Temperature
Out Scale Factor section (below).
If the goal is not met, repeat the Temperature In Scale Factor
procedures.
TEMPERATURE OUT SCALE FACTOR
Repeat as for Temperature In Scale Factor, making the following substitutions:
1. At the Calibration/Configuration Menu:
6 = TEMPERATURE OUT SCALE FACTOR
2. In the screen status display area, read the right column:
COOLING WATER
I/O 22.7 23.2 degC