EasyManuals Logo

Stryker 1188 Series User Manual

Stryker 1188 Series
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
7
Maintenance, Inspection, and Testing
Inspect the device on a continual basis. If a problem is
observed or suspected, the device should be returned
for repair.
Inspect all components for cleanliness. If fluid or tissue
buildup is present, repeat the above cleaning and
disinfection procedures.
Packaging
N/A
Sterilization
After performing the cleaning instructions specified above, perform
one of the following sterilization cycles.
Steam
Note for United States users: For all autoclave-compatible devices,
Stryker recommends using steam sterilization instead of liquid
chemical sterilization.
Steam sterilization is intended only for couplers marked
AUTOCLAVE.
Rapid cooling, orquenching,” the coupler after autoclaving
will result in product damage.
The water used in the autoclave process must meet standards
for clean steam per AAMI ST 79 Appendix M – Steam Quality
requirements.
Sterilizer
Type
“Flash”
Gravity
“Flash”
Pre-vacuum
Gravity Pre-vacuum
Min.
Te m p.
132 – 137°C
(270 – 279°F)
132 – 137°C
(270 – 279°F)
132 – 137°C
(270 – 279°F)
132 – 137°C
(270 – 279°F)
Min. Cycle
Time
10 minutes 3 minutes 10 minutes 3 minutes
Product
Config-
uration
Unwrapped Unwrapped Double
wrapped
Double
wrapped
Drying
Time
8 minutes 8 minutes

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker 1188 Series and is the answer not in the manual?

Stryker 1188 Series Specifications

General IconGeneral
BrandStryker
Model1188 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals