6
Cleaning: Automated
1. Brush
Brush both ends of any lumens a minimum of five
times, using an appropriate bottle brush.
2. Rinse
Rinse the device with treated water at ambient
temperature until there is no visible detergent residue.
Continue to rinse for a minimum of 30 seconds after all
detergent residue has been removed.
Place the device in the washer on an incline to facilitate
drainage.
3. Automated wash
Program the washer using the following parameters:
Disinfection (optional)
1. Disinfect the device by submerging it in a disinfection agent.
Follow the disinfection agent manufacturer’s recommended
concentrations, temperatures, and exposure times.
2. Thoroughly rinse and flush all parts and lumens with running,
demineralized water to remove the disinfectant.
3. Dry all parts with a lint-free towel immediately after rinsing.
Drying
For automated drying, use the drying cycle provided with the
washer/disinfector.
For manual drying, use a lint-free cloth.
Dry any lumens with compressed air.
* If necessary, use pressurized air to aid in drying. Visually inspect each
device for cleanliness.
Phase Recirculation
Time
Water
Te mp er atu re
Detergent Type
and Concentration
(If applicable)
Pre Wash 2 minutes Cold N/A
Enzyme
Wa sh
2 minutes Hot Enzymatic
Detergent
Wash 1 2 minutes Set Point
(66°C)
Regular Detergent
Rinse 1 2 minutes Hot N/A
Dry
Phase
7 minutes 115°C N/A