RReeccoommmmeennddeedd sseeppaarraattiioonn ddiissttaanncceess bbeettwweeeenn ppoorrttaabbllee aanndd mmoobbiillee RRFF ccoommmmuunniiccaattiioonn eeqquuiippmmeenntt aanndd tthhee 22886633
AAIIRR++ ppuummpp
380-390 TETRA 400 1.8 0.3
430-470
GMRS 460;
FRS 460
2.0 0.3
704-787
LTE Band 13, 17
0.2 0.3
800-960
GSM 800/900;
TETRA 800;
iDEN 820;
CDMA 850;
LTE Band 5
2.0 0.3
1,700-1,990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3, 4, 25;
UMTS
2.0 0.3
2,400-2,570
Bluetooth;
WLAN;
802.11 b/g/n;
RFID 2450;
LTE Band 7
2.0 0.3
5,100-5,800
WLAN 802.11 a/n 0.2 0.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NNoottee -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
The 2863 AAIIRR++ pump is suitable for use in the electromagnetic environment specified below. The customer or the user of
the 2863 AAIIRR++ pump should assure that it is used in such an environment.
IImmmmuunniittyy tteesstt
IIEECC 6600660011 tteesstt lleevveell
CCoommpplliiaannccee lleevveell
EElleeccttrroommaaggnneettiicc
eennvviirroonnmmeenntt--gguuiiddaannccee
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
+8 kV contact
+15 kV air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrostatic fast transient/
burst
IEC 61000-4-4
+2 kV for power supply lines
+1 kV for input/ output lines
+2 kV for power supply lines
+1 kV for input/ output lines
Main power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV for input/output lines ±1 kV for input/output lines
Main power quality should
be that of a typical
commercial or hospital
environment.
EN 18 2863-009-001 Rev B.2