GGuuiiddaannccee aanndd mmaannuuffaaccttuurreerr’ss ddeeccllaarraattiioonn -- eelleeccttrroommaaggnneettiicc iimmmmuunniittyy
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
0%U
T
for 0.5 cycle at 0°,
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0%U
T
for 1 cycle
70%U
T
(30% dip in U
T
) for
25/30 cycles
0% U
T
for 250/300 cycles
0%U
T
for 0.5 cycle at 0°,
45°, 90°, 135°, 180°, 225°,
270°, and 315°
0%U
T
for 1 cycle
70%U
T
(30% dip in U
T
) for
25/30 cycles
0% U
T
for 250/300 cycles
Main power quality should
be that of a typical
commercial or hospital
environment. If the user of
the 2863 AAIIRR++ pump
requires continued
operation during power
main interruptions, it is
recommended that the
device be powered from an
uninterrupted power supply
or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NNoottee -- U
T
is the a.c. mains voltage before applications of the test level.