Stryker ProForm Series - Care and maintenance

Stryker ProForm Series
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CCaarree aanndd mmaaiinntteennaannccee
WWAARRNNIINNGG
Do not wash the internal components of this support surface. Discard the support surface if a contamination is found
inside.
Do not immerse the product in cleaning or disinfectant solutions.
Do not allow liquid to pool on the product.
Always inspect support surface covers (top and bottom) for tears, punctures, excessive wear, and misaligned zippers
before each use. If compromised immediately remove the support surface from service.
Always make sure that you wipe each product with clean water and dry each product after you clean. Some cleaning
agents are corrosive in nature and may cause damage to the product. Failure to follow these cleaning instructions may
void your warranty.
CCAAUUTTIIOONN
Do not allow liquid to seep into the zipper area or watershed cover barrier when you clean the underside of the support
surface. Fluids allowed to come in contact with the zipper may leak into the support surface core.
Do not iron, dry-clean, or tumble dry the support surface covers.
Do not power wash the support surface as this may damage the product.
Always dry the support surface covers before you store, add linens, or place a patient on the surface. A dry product
helps to prevent impaired performance.
Do not over expose the covers to higher concentration chemical solutions as these may degrade the covers.
Do not use accelerated hydrogen peroxides or quaternaries that contain glcyol ether >3% as these chemicals may
damage the support surface cover.
Failure to follow manufacturing instructions may also affect useful life of the support surface cover.
The support surface cover is resistant to the following chemical solutions:
Quaternaries (active ingredient - ammonium chloride) that contain less than 3% glycol ether
Phenolic solution (Matar)
Chlorinated bleach solution (10000 ppm) for use with END406 mattress covers
Chlorinated bleach solution (1000 ppm) for use with Nylon mattress covers
70% isopropyl alcohol
Follow hospital protocol for support surface care between patients, to avoid the risk of cross-contamination and infection.
NNoottee -- Chlorinated bleach solution may cause discoloration of the Nylon mattress covers.
2815-009-001 Rev D.1 11 EN