What to do if Stryker Power-PRO XT Medical Equipment SMRT Pak does not provide sufficient power for cot operation?
- AAimee BarrettJul 28, 2025
If your Stryker Medical Equipment's SMRT Pak isn't providing enough power, try these steps: Fully charge the SMRT Pak for two hours. If the charger indicator LED is solid amber, replace the SMRT Pak. Check the voltage at the power terminals on the Pak for 26 VDC minimum immediately after a full charge; if it's less, replace the Pak. Wait one hour, then charge the SMRT Pak again. If the charger indicator LED turns solid green in less than one minute, contact Stryker Technical Support. If the charger indicator LED flashes green for longer than one minute, replace the Pak.