2017/09 F.2 6506-209-001 REV F www.stryker.com
Operations Manual
Power-PRO™ XT
6506

Table of Contents

Other manuals for Stryker Power-PRO XT

Questions and Answers

  • B
    Brenda SmithJul 28, 2025
    What to do if Stryker Power-PRO XT Medical Equipment SMRT Pak does not provide sufficient power for cot operation?
    • A
      Aimee BarrettJul 28, 2025
      If your Stryker Medical Equipment's SMRT Pak isn't providing enough power, try these steps: Fully charge the SMRT Pak for two hours. If the charger indicator LED is solid amber, replace the SMRT Pak. Check the voltage at the power terminals on the Pak for 26 VDC minimum immediately after a full charge; if it's less, replace the Pak. Wait one hour, then charge the SMRT Pak again. If the charger indicator LED turns solid green in less than one minute, contact Stryker Technical Support. If the charger indicator LED flashes green for longer than one minute, replace the Pak.
  • J
    Jermaine MillerAug 7, 2025
    How to troubleshoot Stryker Power-PRO XT Medical Equipment if charger indicator LEDs are not illuminated when the Pak is inserted?
    • J
      John LarsonAug 7, 2025
      If the charger indicator LEDs on your Stryker Medical Equipment are not lighting up when the Pak is inserted, try these steps: Unplug the charger, wait five seconds, and plug it back in. Ensure the charger power LED is lit; if not, consult the section on SMRT charger power LED issues. Make sure the SMRT Pak is fully inserted, listening for a click. Try another SMRT Pak to isolate the problem. If the LEDs still don't light up, replace the charger. If they do light up, replace the original SMRT Pak.
  • J
    Jennifer ReynoldsAug 16, 2025
    What to do if Stryker Power-PRO XT Medical Equipment charger indicates a temperature delay but the Pak is within the normal operating temperature range?
    • B
      Brian RushAug 16, 2025
      If your Stryker Medical Equipment charger shows a temperature delay even when the Pak is at a normal temperature, remove the SMRT Pak and let it cool at room temperature for at least four hours. Then, reinsert the Pak into the charger. If the charger indicator LED continues to flash amber, replace the Pak.

Summary

Warning;Caution;Note Definition

Summary of safety precautions

Pinch points

Identifies potential pinch points and associated hazards during operation.

Mechanical stability

Discusses factors affecting the cot's stability and safe handling.

Introduction

Product description

Details the Stryker Model 6506 Power-PRO XT cot features and components.

Indications for use

Specifies the intended patient population and environments for the Power-PRO XT.

Expected service life

Outlines the expected lifespan of the cot, charger, and battery.

Specifications - Power-PRO

Charger

Details the electrical specifications for the cot's charger unit.

Standards with required options

Outlines required options to meet specific crash-test and safety standards.

Specifications - SMRT

SMRT charger

Provides electrical and physical specifications for the SMRT charger.

SMRT Pak

Details the specifications for the SMRT Pak battery and its enclosure.

Installation

Installing the cot fastener

Provides instructions for installing the Stryker cot fastener system.

Installing the vehicle safety hook

Step-by-step instructions for securely mounting the vehicle safety hook.

Vehicle configuration

Explains vehicle requirements and compatibility for cot installation.

Positioning of the vehicle safety hook, front to back

Instructions for correct front to back placement of the vehicle safety hook.

Operation

Setting cot load height with jog function

How to adjust and set the cot's load height using the jog function.

Operating guidelines

General rules and best practices for safe and effective operation of the cot.

Loading or unloading a cot with the Power-LOAD option

Instructions for using the Power-LOAD system for cot loading and unloading.

Securing the patient with the G-rated restraint straps

Detailed guide on using G-rated restraint straps for patient security.

Charging the SMRT Pak

Instructions for charging the SMRT Pak using the SMRT charger.

Preventive maintenance

Regular inspection and adjustments

Schedule and procedures for routine inspections and adjustments.

Foot end fastener part replacement schedule

Schedule for replacing foot end fastener parts based on call volume.

EMC information

Guidance and manufacturers declaration - electromagnetic emissions

Information on electromagnetic emissions compliance for the product.

Guidance and manufacturers declaration - electromagnetic immunity

Details on electromagnetic immunity compliance for the product.

Warranty

Warranty exclusions and other conditions

Lists conditions that may void the product's warranty.

To obtain parts and service

Information on how to get replacement parts and technical support.

Stryker Power-PRO XT Specifications

General IconGeneral
ModelPower-PRO XT
CategoryMedical Equipment
Weight Capacity700 lbs (318 kg)
Power SourceBattery
Battery TypeSealed lead acid
Frame MaterialAluminum
HeightAdjustable
Minimum Height14 in (36 cm)
Wheels4
Incline CapabilityUnspecified
ControlHandheld
Safety FeaturesSide rails