12 6506-209-001 REV F www.stryker.com
Introduction
Expected service life
Power-PRO has a seven year expected service life under normal use conditions and with appropriate periodic
maintenance.
The SMRT charger has a seven year expected service life under normal use conditions and with appropriate periodic
maintenance.
The SMRT Pak battery has a two year expected service life under normal use conditions.
Contraindications
None known.
Specifications - Power-PRO
Safe working load
Note: Safe working load indicates
the sum of the patient, mattress,
and accessory weight.
700 lb 318 kg
Maximum unassisted lift capacity
1
500 lb 227 kg
Backrest articulation/shock position (standard fowler -
6506-012-003)
0° to 73°/+15°
Backrest articulation/shock position (1865 fowler
option - 6506-012-004)
0° to 75°/+15°
Backrest articulation/shock position (6506-700-013) 6° to 73°/+15°
Overall length/minimum length/width
81 in./63 in./23 in. 206 cm/160 cm/58 cm
Height
2
Adjustable from 14 in. to
41.5 in.
Adjustable from 36 cm to
105 cm
Weight
3
125 lb 57 kg
Caster diameter/width
6 in./2 in. 15 cm/5 cm
Minimum operators required for loading/ unloading an
occupied cot
2
Minimum operators required for loading/ unloading an
unoccupied cot
1
Recommended fastener systems
Model 6370 or 6377 Floor mount type, Model 6371 Wall
mount type, Model 6390 Power-LOAD, Model 6392
Performance-LOAD
Recommended loading height
4
Up to 36 in. Up to 91 cm
Recommended working height (excluding mattress) 15.75 in. 40 cm
Hydraulic oil
Stryker part number 6500-001-293
Power system
Battery 24 VDC NiCd - SMRT power system