www.stryker.com 6506-209-001 REV F 5
Summary of safety precautions
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
WARNING
Always keep your hands clear of the red safety bar pivots when you load, unload, or change the height position of
the cot.
Always use both hands when you transport the cot.
Power-PRO with the Power-LOAD compatibility option operates primarily at these frequencies: 70 - 85 kHz for
inductive charging and 13.56 MHz±7 kHz, Amplitude Modulated (OOK), ERP: -82.37 dBm. The inductive charging
can operate between these frequencies: 70 - 125 kHz. Other equipment may interfere with the Power-LOAD system,
even if that other equipment complies with CISPR emission requirements.
Always install the in-fastener shut-off system in any emergency vehicle that will be used with this cot if using an
antler style cot fastener.
Install the cot fastener by qualified personnel only. Improper installation could result in injury to the patient or
operator.
Always make sure that all cots meet the installation specifications for the Stryker cot fastener system.
Always position the in-fastener shut-off before you place the cot into service.
Do not attempt to operate the cot when it is loaded into a cot fastener.
Always use the in-fastener shut-off to disable the electronic functionality only. Do not use the in-fastener shut-off for
any other purpose.
Do not remove the battery when the cot is active.
Always operate the cot only when all persons are clear of the mechanisms. Entanglement in powered cot
mechanisms can cause serious injury.
Always inspect SMRT Paks for damage before every use.
Do not allow untrained assistants to assist in the operation of the product.
Do not ride on the base of the cot.
Do not transport the cot sideways to avoid the risk of tipping. Always transport the cot in a lowered position, head end
or foot end first to minimize the risk of tipping.
Always conduct patient monitoring when the cot is idle. If you hydraulically raise or lower the product you may
temporarily affect electronic patient monitoring equipment.
Do not leave a patient unattended. Hold the product while a patient is on the product.
Do not apply a wheel lock when a patient is on the product or when you move the product to avoid the risk of tipping.
Do not use siderails as a patient restraint device.
Always transport the cot at a lower height to reduce the risk of a cot tip. If possible, obtain additional assistance or
take an alternate route.
Always avoid high obstacles, such as curbing, steps, or rough terrain to avoid the risk of the product tipping over.
Always use Power-LOAD with the 6085/6086 Performance-PRO XT, 6500/6506 Power-PRO XT, and 6510/6516
Power-PRO IT cots with the Power-LOAD option only. In certain situations, you can use Power-LOAD as a standard
antler for most X-frame cots, but a rail clamp assembly is required for all cots without the Power-LOAD option.
Always make sure that you use a Power-LOAD date of manufacture cot with the Stryker Model 6390 Power-LOAD
system to avoid the risk of injury.
Always support the load of the patient, cot, and accessories after the weight is off of the ground.
Do not pull or lift on the cot safety bar when you unload the cot.
Always lock the head section into place before you operate the cot.
Do not load the cot into a vehicle with the head section retracted when using a cot fastener. The cot may tip or not
connect with the cot fastener.
Do not install or apply a wheel lock on a product with worn wheels that are less than 6 in. diameter.