EasyManua.ls Logo

Stryker Power-PRO XT - Installing a SMRT Pak; Removing a SMRT Pak from the Cot

Stryker Power-PRO XT
87 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60 6506-209-001 REV F www.stryker.com
Operation
Attaching the mattress (Continued)
Note: The mattress is a Type BF applied part.
Installing a SMRT Pak
The SMRT Pak is designed to be compatible with the Power-PRO XT, Power-PRO IT, and Power-PRO TL cots.
To maximize available battery power, only use SMRT Paks that have been charged within the last 48 hours.
To install the SMRT Pak:
1. Align the tabs in the battery enclosure.
2. Push the SMRT Pak into the enclosure until the latch clicks into place.
The SMRT Pak can remain on the Power-PRO cot between service calls for up to one week or until the cot battery LED
indicator flashes amber.
Removing a SMRT Pak from the cot
After the SMRT Pak has been discharged, remove the SMRT Pak from the Power-PRO cot and replace it with a charged
SMRT Pak.
Running the SMRT Pak repeatedly, without rest periods, can increase the temperature within the cells and reduce life.
For example, lifting a heavy patient up and down several times in rapid succession can reduce the SMRT Pak life.
WARNING
Do not remove the battery when the cot is activated.
Do not attempt to open the battery pack for any reason, to avoid the risk of electric shock. If the battery pack case is
cracked or damaged, do not insert it into the charger. Return damaged battery packs to a service center for
recycling.
Always avoid direct contact with a wet battery or battery enclosures. Contact may cause injury to the patient or
operator.
CAUTION
Always remove the battery if the cot is not going to be used for an extended period of time (more than 24 hours).
To remove the SMRT Pak from the cot:
1. Press the red one hand release button (C) or press the battery release button (A) to release the SMRT Pak (B) from
the cot (Figure 49 on page 61).
2. Slide the released SMRT Pak out of the enclosure.

Table of Contents