Stryker System 5 - RF Separation Distances

Stryker System 5
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Recommended separation distances between portable and mobile RF communications equipment and the System 5
handpiece
The System 5 handpiece is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the System 5 handpiece can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the System 5 handpiece as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
150 kHz to 80 MHz
d = [ ]√P
n/a
n/a
n/a
n/a
n/a
80 MHz to 800 MHz
d = [ ]√P
0.12
0.37
1.17
3.70
11.70
800 MHz to 2.5 GHz
d = [ ]√P
0.23
0.74
2.33
7.37
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Separation distance according to frequency of transmitter
m
3.5
V
1
3.5
E
1
7
E
1