PROBLEM
Instrument does not run or oscillates at a
reduced speed making cutting difficult.
Motor runs but blade does not move.
Battery pack becomes unusually hot
during use.
Blade will not fit into the blade retainer or
cannot be secured.
Handpiece has become noisy and vibrates.
Sporadic electrical interference is expe-
rienced.
Troubleshooting Guidelines*
*This product is not field repairable. In case of operating difficulties, all Stryker products must be returned to Stryker Instruments for repair.
For more information, contact your Stryker Instruments sales representative or call Stryker Customer Service at 1-800-253-3210. Outside the
USA, contact your nearest Stryker subsidiary.
CAUSE
Battery pack is discharged.
Battery pack is expended.
Trigger switch is in the SAFE position.
Drivetrain is malfunctioning.
Drivetrain is malfunctioning.
Circuitry is malfunctioning.
Debris is inside the end of the blade
retainer.
Blade is not a Stryker product.
Blade retainer is damaged.
Blade is not a Stryker product.
Drivetrain is malfunctioning.
Electrical noise is present.
ACTION
Recharge the battery in Stryker charger.
Replace the battery pack.
Set the trigger switch to the FAST or STD
position.
Return the handpiece for repair.
Return the handpiece for repair.
Check the battery pack on the Stryker
charger and replace it if indicated. See
Charger Instructions.
Clean with a small brush.
Use a Stryker blade.
Return the handpiece for repair.
Use a Stryker blade.
Return the handpiece for repair.
Turn off all electrical equipment not in use
in the operating room.
Relocate electrical equipment; increase
spatial distance.
Plug operating room equipment into differ-
ent operating room outlets.