*Remain still during measurement.
2) When a pulse is detected, the pressure value moves to the right side of the display. At
this time the
”
v”
mark flashes at the same rate as pulse rate, and the buzzer
sounds when the oulsation is sufficient.
Flashing
urrent Pressure
alue
3) When the measurement is completed and buzzer sounds beep), Systolic pressure is
displayed on the left and Diastolic pressure on the right
SI
e
of the display.4
2-3 seconds later, pressure value and pulse appear alternately on the display.
Pulse rate
*When a measurement is made with insufficient pressure the “Err’ mark appears. Pressurize to a
valve 30-40 mmHg higher than that previously used and perform the measurement again,
5. Push the air release valve to completely release the air from within the cuff.
6. Turn Power OFF
Press the Power switch to turn power OFF. (See Diagram A)
7. Previous Measurement Display
When the Memory switch is pressed, the previous measurement value is shown for
about 5 seconds.
8. Automatic Power OFF Function
If this unit is
lefl
on after measurement an Automatic Power-OFF function turns the
power OFF after approximately 2 min. 30 seconds. To continue measurement press
the power switch turning the unit ON.