Sunbeam 7622 - How to Measure; Powering On and Initializing; Cuff Application and Pressurization; Measurement in Progress Indication

Sunbeam 7622
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
l When power is turned ON, all display marks appear for approximately 1 second.
l When the
“0”
is displayed the unit
IS
ready for measurement. It the
‘7”
mark appears, release air
in the arm cuff bv squeezing it or removtna the hose plug
Approx.
1
second
2. Attach the arm cuff to the left arm for best results.
3. Pressurize the arm cuff with the inflation bulb.
*
Pressurize the arm cuff to approximately 30-40 mmHg above the estimated systolic blood
pressure to be measured
4. When Pressurization is completed, the automatic exhaust mechanism will
gradually reduce arm cuff pressure, and the
“v“
mark will appear indicating that
measurement is in progress.
1) Pressure value appears on the left, and exhaust speed on the right side of the display.