EasyManua.ls Logo

Supco IAQ50 - RH CALIBRATION; 33% Calibration; 75% Calibration

Supco IAQ50
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
Plug the sensor probe into 33% salt
bottle. Hold down and under
normal mode to enter 33% calibration
(Fig.17). “CAL” and calibrating value
(32.7% if at are blinking on the
LCD with current temperature at the left.
Meter is now calibrating, and will finish
in about 60 minutes when “CAL” and
humidity stop blinking. (Fig.18)
o
25 C)
The meter defaults to be calibrated the
humidity with 33% and 75% salt solution.
The ambient condition is recommended
o
to be at 25 C and stable humidity(better
to be close to the calibrating value). To
abort calibration, hold down for more
than 1second at any time.
RH CALIBRATION
%%
OO
Fig. 17
%%
OO
Fig. 18
After 33% calibration, plug the sensor
probe into 75% salt bottle, then press
to enter 75% calibration (Fig.19).
CAUTION:
Do not calibrate the humidity without
the default calibration salt. Otherwise,
it will cause permanent damage.
M /
X
N
AVG
CAL
ESC
CAL
ESC
SETSET
RESETRESET
33% calibration
Magnesium Chloride - MgCL2 • 6 H2O
75% calibration
Sodium Chloride - NaCL

Related product manuals