MAINTENANCE
Anesthesia Machine
The Supera Anesthesia Innovations anesthesia machine should be serviced by a
certied technician. Only certied service companies will have the factory speci-
cations and access to original replacement parts.
Annual Preventative Maintenance Service
Service will include:
Pre–service low and high pressure test. Inspection and replacement of hoses,
gaskets and seals as needed. Measure pressure relief valve resistance. Inspect
unidirectional valves. Inspection of evacuation system. Inspection of oxygen
system. Verication of vaporizer concentration output. Final low and high pres-
sure test.
Contact Supera for a list of certied service companies.
Daily Inspection - Leak Test
Before each use check for pressure leaks in the machine. Also make certain the
waste gas evacuation system is working properly.
Oxygen Concentrator
*See the operators manual that came with the concentrator.
Your machine includes an oxygen concentrator and the lters must be kept clean
for proper operation.
There are two lters to monitor.
1. Coarse lter on the back of the machine. This can be washed with
mild soap and water. DO NOT INSTALL UNTIL 100% DRY
2. Fine particle lter under the course lter. Replace during the annual
PMS of the anesthesia machine. This is a replacement only lter.
Contact Supera for replacement. 877-620-1500
Failure to monitor the condition of the lters will void the warranty.