SVAN 953 USER MANUAL
1-1
1 INTRODUCTION
The SVAN 953 is digital, Type 2 sound level meter along with analyser meeting IEC 61672:2002
standard. The instrument is dedicated to general acoustic measurements, environmental noise
monitoring, occupational health and safety monitoring. Acoustic dose meter function is also available as
an option with the SV 25 dedicated microphone. The SVAN 953, using computational power of its digital
signal processor, can perform real time statistical calculations.
Three acoustic profiles allow parallel measurements with independently defined filters and RMS
detector time constants (e.g. concurrent Impulse, Fast and Slow measurements are possible).
All required weighting filters (e.g.: Z, A, C) are available with this instrument.
Fast USB 1.1 interface (with 12 MHz clock) creates real time link for the application
of the SVAN 953 as a PC front-end. The measurement results can be stored in large 16 MB, non-volatile
memory or external USB Memory Stick (option) and easy downloaded to any PC using USB 1.1 and
SvanPC software. The usage of the RS 232 interface (option) or IrDA interface (option) is also possible.
The SVAN 953 can be used in hard environmental conditions over the whole work day thanks to
replaceable four AA standard or AA rechargeable (i.e. NiMH, the separate charger is required) batteries
and robust, lightweight construction.
The main features of SVAN 953 instrument are as follows:
SVAN 953 as Sound Level Meter
noise measurements (SPL, LEQ, SEL, L
den
, Ltm3, Ltm5, L
Peak
, L
Max
, L
Min
, L
EPd
and statistics)
with Type 2 accuracy in the frequency range 20 Hz ÷ 20 kHz and with 48 kHz sampling rate
parallel IMPULSE, FAST and SLOW detectors for the measurements with A, C or Z filters
one measurement range 26 dB
A RMS
÷ 141 dB
A PEAK
(5 dB margin from noise level) in the SLM mode
General features of SVAN 953
advanced data logger, up to 12 results logged simultaneously from three independent profiles (PEAK
/ MAX / MIN / RMS results from each profile) with time step down to 2 ms
advanced trigger function
USB 1.1 Client (standard) and RS 232, IrDA, USB Host interfaces (options)
powering from four AA standard batteries (operation time - OT > 12 h) or AA rechargeable batteries
(e.g. NiMH - OT > 16 h), SA 17A external battery pack (OT > 24 h), external DC power source (6 V ÷
15 V) or USB interface (500 mA)
handheld, robust case and light weight (only ca 600 grams with batteries, microphone and preamplifier)
Accessories included
7052S - prepolarised ½” condenser ACO PACIFIC’s microphone with nominal sensitivity 20 mV/Pa
SV 12 - microphone preamplifier
SC 16 - USB 1.1 cable
4x AA alkaline batteries
SvanPC for Windows 95/98/NT/2000/XP
Accessories available
SA 17A - external battery pack
SA 22 - windscreen
SA 43 - carrying case for SVAN 95x and accessories
SA 46 - carrying belt-bag for SVAN 94x and SVAN 95x (leather)
SA 47 - carrying bag for SVAN 95x and accessories (fabric material)
SV 25 - dose meter microphone with integrated preamplifier and cable
SV 55 - RS 232 option for the SVAN 955
SvanPC+ for Windows 98/2000/XP
Software options available
SV 953_1 - 1/1 octave analysis option
SV 953_10 - dose meter without the SV 25 microphone