17. Specications
Measurement range : 12.5mm 25mm
Max error (10µm Resolution) : 10µm 10µm (±1 digit)
Max error (1µm Resolution) 3µm 4µm
Max error (0.1µm Resolution) : 1.8µm 2.2µm
Repeatability : 2µm
Weight : 119g 123g
Measurement force (standard) : 0.65-0.9N 0.65-1.15N
Max. speed of travel : 1.7m/s
No. of measurements/ sec : measurement: 10 meas/s MIN/MAX mode: 20 meas/s
Measurement unit : metric/english (Inch)
Maximum Preset (10µm Resolution) : ±9999.99 mm / ±399.9995 IN
Maximum Preset (1µm Resolution) : ±999.999 mm / ±39.99995 IN
Maximum Preset (0.1µm Resolution) : ±99.9999 mm / ±3.999995 IN
Analogue scale value increment : min 0.2 µm - max 1 mm / min 0.01 thou - 0.05 IN
Analogue scale capacity : min ±5 µm - max ±25 mm / min ± 0.25 thou - max ± 1.25 IN
Measurement system : Sylvac inductive system (patented)
Power : 1 x 3V lithium battery, type CR2032, 220mAh
Average battery life : 8’000 hours (
Bluetooth
®
see chapter 9.1)
Data output : RS232/USB compatible
Working temperature (storage) : +5 to +40°C (-10 to +60°C)
Electromagnetic compatibility : as per EN 61326-1
IP rating (in accordance with IEC60529) : IP 54
Fixing and space envelope : Ø8h6 (3/8’’), interchangable M2.5 (4-48-UNF) probe (as per DIN 878)
23
E