10 Description of controls, messages, and indicators
10.1 Laser
emission
control
When pressed into contact with surface (the skin), the laser diodes are activated and starts to emit laser radiation
through the laser aperture. A single audible signal indicates “start of action” and laser emission.
A laser aperture LED is illuminated to indicate emission of laser radiation
When lifted or moved from the surface contact (skin) the laser diodes stop emitting laser radiation through the
laser aperture. Double audible signals indicate “stop of action”, and thus stop of laser emission. When laser
radiation is stopped, the laser aperture LED is switched off
NOTE. If, by any reason, the laser emission is not stopped when the laser apertures are not in contact with the
surface/skin, the operator can use the push-button on the rear end of the device to cut the power to the laser
diodes and thereby stop the emission of laser radiation immediately
10.2 ON/OFF
Switch the device “ON” or “OFF”.
To indicate connection or disconnection. Each position, “ON” or “OFF”, is a stable position.
When the Green LED is switched on, the device is turned “ON”