19.2 Electromagnetic immunity – guidance and declaration
Guidance and manufacturer’s declaration electromagnetic immunity
The PDCare 904 nm Laser by SYMBYX is intended for use in the electromagnetic environment specified below. The customer or the user of this laser
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30 %
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.