QUALITY CONTROL
5-20 Sysmex KX-21 Operator’s Manual -- Revised October 1998
The screens from the start to the end of analysis are as shown below.
(Ready)
FILE No.1 QC
Ready
(Aspirating)
Aspirating
FILE No.1 QC
(Analyzing)
Analyzing
FILE No.1 QC
(Rinsing)
Rinsing
FILE No.1 QC
Figure 5-2-28: L-J
Control Analysis Screen
While the screen is displaying "Aspirating," keep holding the control
blood container in the aforementioned status. If the container is
removed during its display, correct analysis may not be obtained.
CAUTION: Several seconds after the buzzer sounds "beep, beep" and
"Analyzing" appears on the screen, the rinse cup lowers. Remove the
control blood container by that time.
To remove the container, lower it straight down. Take care not to bend
the sample probe.
NOTE: The sample probe is automatically rinsed, so there is no need to wipe it
clean.